Suggestions

DULCOTEST sensors for chlorine dioxide

Graduated measuring ranges 0.01 – 10 mg/l

Chlorine dioxide measurement with DULCOTEST sensors offers precise, reliable, application-specific measured values in real-time for all oxidation and disinfection tasks.

Product description

Reliable online measurement of chlorine dioxide – with DULCOTEST sensors.

Our product line of DULCOTEST chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary sensors provides three sensor product ranges for different applications. Each product range consists of several types for different measuring ranges. To monitor or control clear water, for instance in the disinfection of potable water, there is the tried-and-tested chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary sensor range CDE2. The CDP1 product range is available for rapidly controlling the chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary content in bottle-washing systems. The innovative sensor type CDR1 can be used in polluted waters, e.g. raw water for the production of potable water, cooling water and wastewater.

Your benefits

  • Three application-optimised sensor ranges available
  • Precise, real-time amperometric measurement for efficient process control (short response time)
  • Amperometric measuring eliminates interference caused by turbidity or colouration
  • Integrated or rapid external temperature compensation eliminates interference caused by influence of temperature
  • Diaphragm-covered sensors for reduced dependence on flow, cross-sensitivities, substances in water and film-forming media
  • Diaphragm-covered measuring electrodes embedded in an electrolyte ensure long service life. This maintains optimum measuring conditions regardless of process conditions

Field of application

The following list of applications are general examples of generally suitable applications for this product. Please contact your local sales representative to talk about the applicable use of this product in your specific case / environment.

Product range CDE: potable, industrial and process water without surfactants with slow temperature changes Product range CDP: bottle-washing systems and other applications with addition of surfactants and rapid temperature changes Product range CDR: polluted industrial / process water, water containing surfactants, cooling water, irrigation water, wastewater with little pollution, hot water

    Technical Details

    Sensor type CDE 2-mA

    • Measuring ranges 0.01…10 mg/l
    • Temperature 5…45 °C
    • Pressure max. 1.0 bar
    • pH range 4.0…11.0
    • Response time 120 s
    Sensor type CDP 1-mA
    • Measuring ranges 0.02…2.00 mg/l
    • Temperature 10…45 °C
    • Pressure max. 3.0 bar
    • pH range 5.5…10.5
    • Response time 60 s
    Sensor type CDR 1-mA
    • Measuring ranges 0.01…10.0 mg/l
    • Temperature 1…55 °C
    • Pressure max. 3.0 bar
    • pH range 1.0…10.0
    • Response time 180 s

    Chlorine Dioxide Sensor CDE 2-mA

    Standard sensor for the measurement of chlorine dioxide without cross-sensitivity by free chlorine. For operation on controllers with 4-20 mA input

    Your Benefits

    • Measured variable: Chlorine dioxide, no cross-sensitivity towards free chlorine
    • Diaphragm-covered sensor minimises faults caused by changing flow or ingredients in the water

    Measured variable

    Chlorine dioxide (ClO2)

    Reference method

    DPD1

    pH-range

    4.0…11.0

    Cross sensitivity

    Ozone

    Temperature

    5…45 °C

    Max. pressure

    1.0 bar

    Flow

    DGMa, DLG III: 60...80 l/h
    BAMa: 5...100 l/h (depending on design)

    Supply voltage

    16...24 V DC (2-wire)

    Output signal

    4-20 mA ≈ measuring range, temperature-compensated, uncalibrated, not electrically isolated

    Response time sensor t90

    120 s

    Selectivity

    Chlorine dioxide selective towards free chlorine, chlorite and chlorate

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DGMa, DLG III

    Controllers

    D1C, DAC

    Typical applications

    Uncontaminated drinking water (surfactant-free).

    Resistance to

    Salts, acids, alkalis. Not surfactants

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CDE 2-mA-0.5 ppm

    0.01…0.5 mg/l

    792930

    CDE 2-mA-2 ppm

    0.02…2.0 mg/l

    792929

    CDE 2-mA-10 ppm

    0.10…10.0 mg/l

    792928

    Chlorine dioxide sensors complete with 100 ml of electrolyte

    Note: a mounting kit (order no. 815079) is required for initial fitting of the chlorine dioxide sensors in the in-line probe housing DLG III.

    Chlorine Dioxide Sensor CDP 1-mA

    Sensor for the measurement of chlorine dioxide with a fast response time, for example in bottle-washing systems. For operation on controllers with 4-20 mA input

    Your Benefits

    • Measured variable: Chlorine dioxide without interference caused by surfactants
    • Diaphragm-covered sensor minimises faults caused by changing flow or ingredients in the water
    • Fast response time through open-pored diaphragm and external temperature measurement

    Field of Application

    • Bottle washing system

    Measured variable

    Chlorine dioxide (ClO2)

    Reference method

    DPD1

    pH-range

    5.5…10.5

    Cross sensitivity

    Ozone, chlorine

    Temperature

    10…45 °C

    Max. pressure

    3.0 bar

    Flow

    DGMa, DLG III: 40...60 l/h
    BAMa: 5...100 l/h (depending on design)

    Supply voltage

    16...24 V DC (2-wire)

    Output signal

    4-20 mA ≈ measuring range, not temperature-compensated, uncalibrated, not electrically isolated

    Temperature measurement

    Separate temperature measurement needed for compensation

    Response time sensor t90

    60 s

    Selectivity

    Chlorine dioxide as against chlorite and chlorate

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    We would recommend installing the sensor together with a Pt 100 temperature sensor in fittings BAMa, DGMa, DLG III

    Controllers

    D1C and DAC with automatic temperature correction only

    Typical applications

    Process water containing surfactants (bottle washing machines).

    Resistance to

    Surfactants, slight films of dirt

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CDP 1-mA-2 ppm

    0.02…2.0 mg/l

    1002149

    Chlorine dioxide sensors complete with 100 ml of electrolyte

    Note: a mounting kit (order no. 815079) is required for initial fitting of the chlorine dioxide sensors in the in-line probe housing DLG III.

    Chlorine Dioxide Sensor CDR 1-mA

    Sensor for the measurement of chlorine dioxide for all kinds of water, including hot and contaminated water. Without cross-sensitivity by free chlorine. For operation on controllers with 4-20 mA input

    Your Benefits

    • Measured variable: Chlorine dioxide, without cross-sensitivity towards free chlorine
    • Diaphragm-covered sensor minimises faults caused by changing flow or ingredients in the water
    • Resistance to films of dirt by pore-free diaphragm
    • Operating temperature up to 60 °C (short term) by appropriate sensor materials

    Measured variable

    Chlorine dioxide (ClO2)

    Reference method

    DPD1

    pH-range

    1.0…10.0

    Cross sensitivity

    Ozone

    Temperature

    1…55 °C

    Max. pressure

    3.0 bar

    Flow

    DGMa, DLG III: 30...60 l/h
    BAMa: 5...100 l/h (depending on design)

    Supply voltage

    16…24 V DC

    Output signal

    4-20 mA temperature-compensated, uncalibrated, not electrically isolated

    Response time sensor t90

    3 min.

    Selectivity

    Chlorite

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DGMa, DLG III

    Controllers

    D1C, DAC

    Typical applications

    Contaminated industrial, process water, containing surfactants, cooling water, irrigation water, slightly contaminated wastewater, warm water.

    Resistance to

    Surfactants, slight films of dirt, water-soluble chemicals, solids/dirt, biofilms

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CDR 1-mA-0.5 ppm

    0.01…0.5 mg/l

    1033762

    CDR 1-mA-2 ppm

    0.02…2.0 mg/l

    1033393

    CDR 1-mA-10 ppm

    0.10…10.0 mg/l

    1033404

    Note: a mounting kit (order no. 815079) is required for initial fitting of the chlorine dioxide sensors in the in-line probe housing DLG III.

    Chlorine Dioxide Sensor CDR 1-CAN

    Sensor for the measurement of chlorine dioxide for all kinds of water, including hot and contaminated water. Without cross-sensitivity by free chlorine. For operation on controllers with 4-20 mA input

    Your Benefits

    • Measured variable: Chlorine dioxide, without cross-sensitivity to free chlorine
    • Diaphragm-covered sensor minimises faults caused by changing flow or ingredients in the water
    • Resistance to films of dirt by pore-free diaphragm
    • Operating temperature up to 60 °C (short term) by appropriate sensor materials
    • Operation on the CAN-bus with all the associated benefits

    Measured variable

    Chlorine dioxide (ClO2)

    Reference method

    DPD1

    pH-range

    1.0…10.0

    Cross sensitivity

    Ozone

    Temperature

    5…45 °C

    Max. pressure

    1.0 bar

    Flow

    DGMa, DLG III: 30...100 l/h
    BAMa: 5...60 l/h (depending on design)

    Supply voltage

    Via CAN-interface (11 – 30 V)

    Output signal

    Uncalibrated, temperature compensated, electrically isolated

    Response time sensor t90

    3 min.

    Selectivity

    Chlorite

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DGMa, DLG III

    Controllers

    DULCOMARIN

    Typical applications

    Contaminated industrial, process water, containing surfactants, cooling water, irrigation water, slightly contaminated wastewater.

    Resistance to

    Surfactants, water-soluble pollutants, solids/dirt, biofilms

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CDR 1-CAN-10 ppm

    0.01…10.0 mg/l

    1041155

    Complete with 100 ml of electrolyte, connecting cable - CAN M12 5-pin 0.5 m, T-distributor M12 5-pin CAN

    Downloads for DULCOTEST sensors for chlorine dioxide

    More in the Download Centre

    0 Results
    Filter by document type
    Application Example / Reference (0)
    Brochure / Flyer (0)
    Catalogue (0)
    Certificate / Declaration (0)
    Data Sheet (0)
    Drawing / Dimensional Drawing / 3D (0)
    Exploded View / Spare Parts (0)
    Material Safety Data Sheet (0)
    Operating Instructions (0)
    Others (0)
    Press Release (0)
    Publication / Article (0)
    Software (0)
    Wiring Diagram (0)

    Selected filters:

    Matching downloads

    Sort by Document type
    Load more matches ...
    No more matches available!

    Unfortunately, we did not find an appropriate match.

    All words written correctly?
    Search again with fewer or different search terms?

    Please try again later

    Service Unavailable

    The server is temporarily unable to service your request due to maintenance downtime or capacity problems. Please try again later.

    Overview

    Reliable online measurement of chlorine dioxide – with DULCOTEST sensors.

    Our product line of DULCOTEST chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary sensors provides three sensor product ranges for different applications. Each product range consists of several types for different measuring ranges. To monitor or control clear water, for instance in the disinfection of potable water, there is the tried-and-tested chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary sensor range CDE2. The CDP1 product range is available for rapidly controlling the chlorine dioxide Chlorine dioxide is an exceptionally reactive gas, which is not stored due to its instability ... more in the glossary content in bottle-washing systems. The innovative sensor type CDR1 can be used in polluted waters, e.g. raw water for the production of potable water, cooling water and wastewater.

    Your benefits

    • Three application-optimised sensor ranges available
    • Precise, real-time amperometric measurement for efficient process control (short response time)
    • Amperometric measuring eliminates interference caused by turbidity or colouration
    • Integrated or rapid external temperature compensation eliminates interference caused by influence of temperature
    • Diaphragm-covered sensors for reduced dependence on flow, cross-sensitivities, substances in water and film-forming media
    • Diaphragm-covered measuring electrodes embedded in an electrolyte ensure long service life. This maintains optimum measuring conditions regardless of process conditions

    Application

    Technical Data

    Downloads

    Contact us

    Yestin Ng

    Managing Director, ProMinent Fluid Controls (M) Sdn Bhd

    »Please send us your request! We will contact you shortly.«

    +603-8068 2578  Send request   

    Matching Products

    The solenoid-driven diaphragm metering pump gamma incorporates a wealth of eX cellent ingenuity! With integrated pressure measurement, it ensures the smooth running of your metering process. The gamma/ X is ideal for all metering work involving liquid media.

    more

    Do you wish for a simple controller for water analysis? One that is easy to operate and with which you can freely select between all common measured variables per channel? There is one: our all-rounder DULCOMETER diaLog DACb! What is more, it is Ethernet-/LAN-capable and can be ideally integrated into existing networks.

    more

    Photometers measure nearly all disinfectants and the pH value based on the photometric principle. They are portable, compact and make safe, simple measurement possible.

    more

    Controller AEGIS II continuously measures and controls the conductivity and biocide concentration to keep pipework and heat exchangers clean.

    more

    Chlorine dioxide system, which continuously produces ClO2 according to the acid/chlorite method with diluted chemicals. Extremely simple operation, clear construction, analogue control, manual control or via contacts.

    more

    Chlorine dioxide system for production of a chlorine-free chlorine dioxide solution, especially suitable for multiple points of injection. Bello Zon CDLb produces ClO2 discontinuously using the acid/chlorite process with diluted chemicals.

    more

    Chlorine dioxide system for the metering of chlorine dioxide with diluted starting chemicals. The certified yield guarantees efficient chlorine dioxide production. Bello Zon CDVd can be easily and safely integrated into any water treatment process.

    more

    Chlorine dioxide system for the metering of chlorine dioxide with concentrated starting chemicals. The proven safety concept protects people and the environment. The certified yield guarantees efficient chlorine dioxide production. Bello Zon CDKd can be easily and safely integrated into any water treatment process.

    more

    All-purpose solenoid-driven metering pump for metering liquid media in water treatment and chemical processes: Solenoid-driven metering pump beta. Cost-effective, overload-proof and adaptable to existing signal transducers.

    more

    The gamma/ XL is a smart, connectible solenoid-driven metering pump that is setting new standards in terms of productivity, reliability and cost-effectiveness.

    more

    Monitoring and treatment of potable and similar types of water with DULCOTROL Potable water/F&B – the compact measuring and control system specially designed for water treatment in waterworks and in the food and beverage industry.

    more

    Chlorite measurements with DULCOTEST sensors in disinfection processes involving chlorine dioxide – high process reliability guaranteed.

    more

    ProMinent Newsletter - receive latest news, customized for you! subscribe now